Product description
Features
complete probe set, for measurements up to 250MHz
attenuation: 1:1 or 1:10
safety: conform to IEC-1010 cat II
Specifications
position x1:
attenuation: 1:1
bandwidth: DC to 10MHz
rise time: 35ns
input resistance: 1Mohm (input resistance scope)
input capacitance: cap. scope 46pF
working voltage: 600Vdc incl. peak AC derating with frequency
position REF: tip disconnected, scope input at ground
position x10:
attenuation: 1:10
bandwidth: DC to 250MHz
rise time: 1.4ns
input resistance: 10Mohm (if scope input = 1Mohm)
input capacitance: ± 11pF
working voltage: 600Vdc incl. peak AC derating with frequency
compensation range: 10 to 35pF
cable length: 1.2m
complete probe set, for measurements up to 250MHz
attenuation: 1:1 or 1:10
safety: conform to IEC-1010 cat II
Specifications
position x1:
attenuation: 1:1
bandwidth: DC to 10MHz
rise time: 35ns
input resistance: 1Mohm (input resistance scope)
input capacitance: cap. scope 46pF
working voltage: 600Vdc incl. peak AC derating with frequency
position REF: tip disconnected, scope input at ground
position x10:
attenuation: 1:10
bandwidth: DC to 250MHz
rise time: 1.4ns
input resistance: 10Mohm (if scope input = 1Mohm)
input capacitance: ± 11pF
working voltage: 600Vdc incl. peak AC derating with frequency
compensation range: 10 to 35pF
cable length: 1.2m